Wall-jet conductivity detector for microchip capillary electrophoresis

MPS-Authors
/persons/resource/persons4056

Muck,  A.
Research Group Mass Spectrometry, MPI for Chemical Ecology, Max Planck Society;

External Resource
No external resources are shared
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Wang, J., Chen, G., & Muck, A. (2009). Wall-jet conductivity detector for microchip capillary electrophoresis. Talanta, 78(1), 207-211. doi:10.1016/j.talanta.2008.10.055.


Cite as: https://hdl.handle.net/11858/00-001M-0000-0012-B4DA-B
Abstract
A new end-column 'hybrid' contactless conductivity detector for microchip capillary electrophoresis (CE) was developed. It is based on a "hybrid" arrangement where the receiving electrode is insulated by a thin layer of insulator and placed in the bulk