Analytical Modeling of the Coupling within a Human Head Surface Loop Transmit Phased Array at Ultra-High Fields

MPS-Authors
/persons/resource/persons133464

Avdievich,  NI
Research Group MR Spectroscopy and Ultra-High Field Methodology, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons192811

Pfrommer,  A
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Research Group MR Spectroscopy and Ultra-High Field Methodology, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons192635

Giapitzakis,  IA
Research Group MR Spectroscopy and Ultra-High Field Methodology, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

/persons/resource/persons84402

Henning,  A
Max Planck Institute for Biological Cybernetics, Max Planck Society;
Research Group MR Spectroscopy and Ultra-High Field Methodology, Max Planck Institute for Biological Cybernetics, Max Planck Society;
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

External Resource

Link
(Any fulltext)

Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)
There are no public fulltexts stored in PuRe
Supplementary Material (public)
There is no public supplementary material available
Citation

Avdievich, N., Pfrommer, A., Giapitzakis, I., & Henning, A. (2016). Analytical Modeling of the Coupling within a Human Head Surface Loop Transmit Phased Array at Ultra-High Fields. Poster presented at 24th Annual Meeting and Exhibition of the International Society for Magnetic Resonance in Medicine (ISMRM 2016), Singapore.


Cite as: https://hdl.handle.net/21.11116/0000-0000-7BA5-C
Abstract
Decoupling of multi-channel ultra-high field (>7T) transmit and transceiver arrays is a major issue. Analytical modeling of the coupling can facilitate the array optimization. We developed an analytical model describing the impedance matrix for two rectangular loops placed on a cylindrical surface and mimicking the human head array geometry. The developed model was comprehensively validated and allows for the optimization of the geometry and positioning of the loops. The latter enabled simultaneous cancellation of resistive and inductive coupling without additional decoupling circuits. The resulting overlapped array element arrangement improves both transmit and receive performance in comparison to conventional gapped arrays.