The influence of flow and motion in MRI of diffusion using a modified CE-FAST sequence.

MPG-Autoren
/persons/resource/persons15516

Merboldt,  K. D.
Research Group of Biomedical NMR, MPI for Biophysical Chemistry, Max Planck Society;

/persons/resource/persons15172

Hänicke,  W.
Research Group of Biomedical NMR, MPI for Biophysical Chemistry, Max Planck Society;

/persons/resource/persons15162

Gyngell,  M.L.
Research Group of Biomedical NMR, MPI for Biophysical Chemistry, Max Planck Society;

/persons/resource/persons15082

Frahm,  J.
Research Group of Biomedical NMR, MPI for Biophysical Chemistry, Max Planck Society;

/persons/resource/persons14899

Bruhn,  H.
Research Group of Biomedical NMR, MPI for Biophysical Chemistry, Max Planck Society;

Externe Ressourcen
Es sind keine externen Ressourcen hinterlegt
Volltexte (beschränkter Zugriff)
Für Ihren IP-Bereich sind aktuell keine Volltexte freigegeben.
Volltexte (frei zugänglich)

2376697.pdf
(Verlagsversion), 650KB

Ergänzendes Material (frei zugänglich)
Es sind keine frei zugänglichen Ergänzenden Materialien verfügbar
Zitation

Merboldt, K. D., Hänicke, W., Gyngell, M., Frahm, J., & Bruhn, H. (1989). The influence of flow and motion in MRI of diffusion using a modified CE-FAST sequence. Magnetic Resonance in Medicine, 12(2), 198-208. doi:10.1002/mrm.1910120206.


Zitierlink: https://hdl.handle.net/11858/00-001M-0000-002C-29A0-5
Zusammenfassung
Severe motion and flow artifacts are a problem in MRI of diffusion in vivo due to the application of strong magnetic field gradients. Here it is shown that image artifacts can be removed by using a modified fast-scan MRI sequence (CE-FAST) in conjunction with averaging of diffusion-weighted images. In phantom studies slow (coherent) flow (<1 mm s-1) in the presence of strong diffusion gradients is shown to cause signal losses in diffusion-weighted images that depend on the relative orientations of the flow direction and the diffusion gradient. On the other hand, pulsatile motions of macroscopic dimensions (e.g. 1 mm, 1 Hz, in-plane) lead to smearing and ghosting of signal intensities along the phase-encoding direction of the images. In both phantoms and rabbit brains in vivo motion artifacts were found to be reducible by averaging 8-16 images. Unfortunately. the resulting image contrast no longer represents a “true” diffusion contrast but is affected by additional signal losses due to motion averaging. All experiments were performed on a 40-cm-borc 2.35-T Bruker Medspee system.