Krypton assay in xenon at the ppq level using a gas chromatographic system combined with a mass spectrometer

MPS-Authors
http://pubman.mpdl.mpg.de/cone/persons/resource/persons30766

Lindemann,  Sebastian
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

http://pubman.mpdl.mpg.de/cone/persons/resource/persons31050

Simgen,  Hardy
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Locator
There are no locators available
Fulltext (public)

1308.4806.pdf
(Preprint), 459KB

Supplementary Material (public)
There is no public supplementary material available
Citation

Lindemann, S., & Simgen, H. (2014). Krypton assay in xenon at the ppq level using a gas chromatographic system combined with a mass spectrometer. The European Physical Journal C: Particles and Fields, 74: 2746. doi:10.1140/epjc/s10052-014-2746-1.


Cite as: http://hdl.handle.net/11858/00-001M-0000-0014-73AB-C
Abstract
We have developed a new method to measure krypton traces in xenon at so far unprecedented low concentrations. This is a mandatory task for many near-future low-background particle physics detectors. Our system is based on a cryogenic gas chromatographic krypton/xenon separation and a subsequent mass spectroscopic krypton quantification. We prove this system to reach a detection limit of 8 ppq (parts per quadrillion) and present results of distilled xenon with krypton concentrations below 1 ppt.