Echo Planar Imaging at 7T with 3rd Order Slice-Wise Dynamic Shim Update (DSU) and Full Eddy-Current Compensation (ECC)

MPS-Authors
http://pubman.mpdl.mpg.de/cone/persons/resource/persons84402

Henning,  A
Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society;

Locator
There are no locators available
Fulltext (public)
There are no public fulltexts available
Supplementary Material (public)
There is no public supplementary material available
Citation

Fillmer, A., Vannesjö SJ, Pavan M, Pruessmann KP, Boesiger, P., & Henning, A. (2012). Echo Planar Imaging at 7T with 3rd Order Slice-Wise Dynamic Shim Update (DSU) and Full Eddy-Current Compensation (ECC). Poster presented at 20th Annual Meeting and Exhibition of the International Society for Magnetic Resonance in Medicine (ISMRM 2012), Melbourne, Australia.


Cite as: http://hdl.handle.net/11858/00-001M-0000-0013-B79C-1
Abstract
The transition to ultra high field strengths yields advantages as increased SNR and higher BOLD contrast, but at the expense of decreased B1 and B0 homogeneity. This work presents the application of Dynamic Shim Updating (DSU) with full 3rd Order eddy-current compensation to Echo Planar Images, which are the basis for fMRI. Due to a highly accurate calibration of the pre-emphasis very fast switching of shim terms during the sequence is possible, without extending the scan time. The presented results exhibit a significant gain in image quality.