Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment

MPS-Authors
/persons/resource/persons30280

Bauer,  C.
Division Prof. Dr. Werner Hofmann, MPI for Nuclear Physics, Max Planck Society;

Borger,  E.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Hofacker,  R.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

/persons/resource/persons30647

Kaether,  F.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

/persons/resource/persons30743

Langbrandtner,  C.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

/persons/resource/persons30768

Lindner,  Manfred
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Reissfelder,  M.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

/persons/resource/persons31005

Schoenert,  S.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

External Resource
Fulltext (restricted access)
There are currently no full texts shared for your IP range.
Fulltext (public)

1104.0758
(Preprint), 4KB

Supplementary Material (public)
There is no public supplementary material available
Citation

Bauer, C., Borger, E., Hofacker, R., Jänner, K., Kaether, F., Langbrandtner, C., et al. (2011). Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment. Journal of Instrumentation, 6(06): P06008, pp. 1-17. doi:10.1088/1748-0221/6/06/P06008.


Cite as: https://hdl.handle.net/11858/00-001M-0000-000F-06B4-8
Abstract
The hemispherical 10" photomultiplier tube (PMT) R7081 from Hamamatsu Photonics K.K. (HPK) is used in various experiments in particle and astroparticle physics. We describe the test and calibration of 474 PMTs for the reactor antineutrino experiment Double Chooz. The unique test setup at Max-Planck-Institut f\"ur Kernphysik Heidelberg (MPIK) allows one to calibrate 30 PMTs simultaneously and to characterize the single photo electron response, transit time spread, linear behaviour and saturation effects, photon detection efficiency and high voltage calibration.