de.mpg.escidoc.pubman.appbase.FacesBean
Deutsch
 
Hilfe Wegweiser Impressum Kontakt Einloggen
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT

Freigegeben

Zeitschriftenartikel

Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment

MPG-Autoren
http://pubman.mpdl.mpg.de/cone/persons/resource/persons30280

Bauer,  C.
Division Prof. Dr. Werner Hofmann, MPI for Nuclear Physics, Max Planck Society;

Borger,  E.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Hofacker,  R.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

http://pubman.mpdl.mpg.de/cone/persons/resource/persons30647

Kaether,  F.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

http://pubman.mpdl.mpg.de/cone/persons/resource/persons30743

Langbrandtner,  C.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

http://pubman.mpdl.mpg.de/cone/persons/resource/persons30768

Lindner,  Manfred
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Reissfelder,  M.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

http://pubman.mpdl.mpg.de/cone/persons/resource/persons31005

Schoenert,  S.
Division Prof. Dr. Manfred Lindner, MPI for Nuclear Physics, Max Planck Society;

Externe Ressourcen
Volltexte (frei zugänglich)

1104.0758
(Preprint), 4KB

Ergänzendes Material (frei zugänglich)
Es sind keine frei zugänglichen Ergänzenden Materialien verfügbar
Zitation

Bauer, C., Borger, E., Hofacker, R., Jänner, K., Kaether, F., Langbrandtner, C., et al. (2011). Qualification Tests of 474 Photomultiplier Tubes for the Inner Detector of the Double Chooz Experiment. Journal of Instrumentation, 6(06): P06008, pp. 1-17. doi:10.1088/1748-0221/6/06/P06008.


Zitierlink: http://hdl.handle.net/11858/00-001M-0000-000F-06B4-8
Zusammenfassung
The hemispherical 10" photomultiplier tube (PMT) R7081 from Hamamatsu Photonics K.K. (HPK) is used in various experiments in particle and astroparticle physics. We describe the test and calibration of 474 PMTs for the reactor antineutrino experiment Double Chooz. The unique test setup at Max-Planck-Institut f\"ur Kernphysik Heidelberg (MPIK) allows one to calibrate 30 PMTs simultaneously and to characterize the single photo electron response, transit time spread, linear behaviour and saturation effects, photon detection efficiency and high voltage calibration.