  Laser-Frequency Stabilization via a Quasimonolithic Mach-Zehnder Interferometer with Arms of Unequal Length and Balanced dc Readout

Gerberding, O., Isleif, K.-S., Mehmet, M., Danzmann, K., & Heinzel, G. (2017). Laser-Frequency Stabilization via a Quasimonolithic Mach-Zehnder Interferometer with Arms of Unequal Length and Balanced dc Readout. Physical Review Applied, 7: 024027. doi:10.1103/PhysRevApplied.7.024027.

Item is

Basisdaten

einblenden: ausblenden:
Datensatz-Permalink: http://hdl.handle.net/11858/00-001M-0000-002C-3A04-F Versions-Permalink: http://hdl.handle.net/11858/00-001M-0000-002E-2FB7-6
Genre: Zeitschriftenartikel

Dateien

einblenden: Dateien
ausblenden: Dateien
:
1610.09684.pdf (Preprint), 745KB
Beschreibung:
File downloaded from arXiv at 2017-01-12 16:09
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
:
PRA.7.024027.pdf (Verlagsversion), 866KB
 
Datei-Permalink:
-
Beschreibung:
-
Sichtbarkeit:
Eingeschränkt
MIME-Typ / Prüfsumme:
application/pdf
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
American Physical Society
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Gerberding, Oliver1, Autor              
Isleif, Katharina-Sophie, Autor
Mehmet, Moritz1, Autor              
Danzmann, Karsten1, Autor              
Heinzel, Gerhard1, Autor              
Affiliations:
1Laser Interferometry & Gravitational Wave Astronomy, AEI-Hannover, MPI for Gravitational Physics, Max Planck Society, escidoc:24010              

Inhalt

einblenden:
ausblenden:
Schlagwörter: Physics, Instrumentation and Detectors, physics.ins-det
 Zusammenfassung: Low frequency high precision laser interferometry is subject to excess laser frequency noise coupling via arm-length differences which is commonly mitigated by locking the frequency to a stable reference system. This is crucial to achieve picometer level sensitivities in the 0.1 mHz to 1 Hz regime, where laser frequency noise is usually high and couples into the measurement phase via arm-length mismatches in the interferometers. Here we describe the results achieved by frequency stabilising an external cavity diode laser to a quasi-monolithic unequal arm-length Mach-Zehnder interferometer read out at mid-fringe via balanced detection. This stabilisation scheme has been found to be an elegant solution combining a minimal number of optical components, no additional laser modulations and relatively low frequency noise levels. The Mach-Zehnder interferometer has been designed and constructed to minimise the influence of thermal couplings and to reduce undesired stray light using the optical simulation tool IfoCAD. We achieve frequency noise levels corresponding to LISA-like (laser interferometer space antenna) displacement sensitivities below 1 pm/$\sqrt{\textrm{Hz}}$ and are able to demonstrate the LISA frequency pre-stabilisation requirement of 300 Hz/$\sqrt{\textrm{Hz}}$ down to frequencies of 100 mHz by beating the stabilised laser with an Iodine locked reference.

Details

einblenden:
ausblenden:
Sprache(n):
 Datum: 2016-10-302017
 Publikationsstatus: Im Druck publiziert
 Seiten: 6 pages, 6 figures, submitted to Phyiscal Review Applied
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: arXiv: 1610.09684
URI: http://arxiv.org/abs/1610.09684
DOI: 10.1103/PhysRevApplied.7.024027
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Physical Review Applied
  Kurztitel : Phys. Rev. Appl.
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: College Park, Md. [u.a.] : American Physical Society
Seiten: - Band / Heft: 7 Artikelnummer: 024027 Start- / Endseite: - Identifikator: ISSN: 2331-7019
CoNE: http://pubman.mpdl.mpg.de/cone/journals/resource/2331-7019