  Weighing of biomolecules, single cells and single nanoparticles in fluid.

Burg, T. P., Godin, M., Shen, W., Carlson, G., Foster, J. S., Babcock, K., et al. (2007). Weighing of biomolecules, single cells and single nanoparticles in fluid. Nature, 446, 1066-1069. doi:10.1038/nature05741.

Item is

Files

show Files
hide Files
:
1851242.pdf (Publisher version), 2MB
Name:
1851242.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-
:
1851242_Supplement_1.pdf (Supplementary material), 256KB
Name:
1851242_Supplement_1.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show
hide
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Burg, T. P.1, Author           
Godin, M., Author
Shen, W., Author
Carlson, G., Author
Foster, J. S., Author
Babcock, K., Author
Manalis, S. R., Author
Affiliations:
1Research Group of Biological Micro- and Nanotechnology, MPI for biophysical chemistry, Max Planck Society, ou_578602              

Content

show
hide
Free keywords: -
 Abstract: Nanomechanical resonators enable the measurement of mass with extraordinary sensitivity1, 2, 3, 4, 5, 6, 7. Previously, samples as light as 7 zeptograms (1 zg = 10-21 g) have been weighed in vacuum, and proton-level resolution seems to be within reach8. Resolving small mass changes requires the resonator to be light and to ring at a very pure tone—that is, with a high quality factor9. In solution, viscosity severely degrades both of these characteristics, thus preventing many applications in nanotechnology and the life sciences where fluid is required10. Although the resonant structure can be designed to minimize viscous loss, resolution is still substantially degraded when compared to measurements made in air or vacuum11, 12, 13, 14. An entirely different approach eliminates viscous damping by placing the solution inside a hollow resonator that is surrounded by vacuum15, 16. Here we demonstrate that suspended microchannel resonators can weigh single nanoparticles, single bacterial cells and sub-monolayers of adsorbed proteins in water with sub-femtogram resolution (1 Hz bandwidth). Central to these results is our observation that viscous loss due to the fluid is negligible compared to the intrinsic damping of our silicon crystal resonator. The combination of the low resonator mass (100 ng) and high quality factor (15,000) enables an improvement in mass resolution of six orders of magnitude over a high-end commercial quartz crystal microbalance17. This gives access to intriguing applications, such as mass-based flow cytometry, the direct detection of pathogens, or the non-optical sizing and mass density measurement of colloidal particles.

Details

show
hide
Language(s): eng - English
 Dates: 2007-04-26
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1038/nature05741
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Nature
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 446 Sequence Number: - Start / End Page: 1066 - 1069 Identifier: -