  T1 quantification with inversion recovery TrueFISP

Scheffler, K. (2001). T1 quantification with inversion recovery TrueFISP. Magnetic Resonance in Medicine, 45(4), 720-723. doi:10.1002/mrm.1097.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Scheffler, K1, Author           
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: A snapshot FLASH sequence can be used to acquire the time course of longitudinal magnetization during its recovery after a single inversion pulse. However, excitation pulses disturb the exponential recovery of longitudinal magnetization and may produce systematic errors in T1 estimations. In this context the possibility of using the TrueFISP sequence to detect the recovery of longitudinal magnetization for quantitative T1 measurements was examined. Experiments were performed on different Gd-doped water phantoms and on humans. T1 values derived from inversion recovery TrueFISP were in excellent agreement with the single-point method even for flip angles up to 50°. In terms of T1 accuracy and SNR, the proposed method seems to be superior to the conventional inversion recovery snapshot FLASH technique.

Details

show
hide
Language(s):
 Dates: 2001-04
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: URI: http://onlinelibrary.wiley.com/doi/10.1002/mrm.1097/pdf
DOI: 10.1002/mrm.1097
BibTex Citekey: SchefflerH2001
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 45 (4) Sequence Number: - Start / End Page: 720 - 723 Identifier: -