English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
 
 
DownloadE-Mail
  A polymer-based MR-compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA)

Mekle, R., Hofmann E, Scheffler, K., & Bilecen, D. (2006). A polymer-based MR-compatible guidewire: A study to explore new prospects for interventional peripheral magnetic resonance angiography (ipMRA). Journal of Magnetic Resonance Imaging, 23(2), 145-155. doi:10.1002/jmri.20486.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Mekle, R, Author
Hofmann E, Scheffler, K1, Author           
Bilecen, D, Author
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: Purpose To introduce a newly developed polymer-based and magnetic resonance (MR)-compatible guidewire and to explore its capabilities with respect to interventional peripheral magnetic resonance angiography (ipMRA) in a flow phantom. Materials and Methods The guidewire is based on a polyetheretherketone (PEEK) polymer core, and small iron particles are embedded in its coating. A passive device tracking technique was designed utilizing a susceptibility artifact induced by the wire in images acquired with a balanced steady-state free precession (b-SSFP) sequence using small flip angles. The position of the guidewire tip was determined from image intensity maxima and overlayed onto a roadmap in near real-time. Guidewire tracking and balloon angioplasty of an artificial stenosis were attempted in two configurations of a flow phantom. Results Successful passive guidewire tracking was performed for all phantom configurations. Robustness and accuracy of the tracking technique were sufficient for phantom studies. A balloon catheter was placed into the stenosis using the guidewire under complete MR guidance, and subsequent balloon angioplasty yielded improved flow conditions. Conclusion The new guidewire is well-suited for clinical application due to an absence of the risk of core fracture and its atraumatic flexible tip. It opens novel prospects for the realization of ipMRA in humans that need to be explored in further studies.

Details

show
hide
Language(s):
 Dates: 2006-02
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: URI: http://onlinelibrary.wiley.com/doi/10.1002/jmri.20486/pdf
DOI: 10.1002/jmri.20486
BibTex Citekey: MekleHSB2006
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Journal of Magnetic Resonance Imaging
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 23 (2) Sequence Number: - Start / End Page: 145 - 155 Identifier: -