Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

 
 
DownloadE-Mail
  Automatic slice positioning (ASP) for passive real-time tracking of interventional devices using projection-reconstruction imaging with echo-dephasing (PRIDE)

Patil, S., Bieri, O., & Scheffler, K. (2009). Automatic slice positioning (ASP) for passive real-time tracking of interventional devices using projection-reconstruction imaging with echo-dephasing (PRIDE). Magnetic Resonance in Medicine, 62(4), 935-942. doi:10.1002/mrm.22080.

Item is

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Patil, S, Autor
Bieri, O, Autor
Scheffler, K1, Autor           
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: A novel and fast approach for passive real-time tracking of interventional devices using paramagnetic markers, termed “projection-reconstruction imaging with echo-dephasing” (PRIDE) is presented. PRIDE is based on the acquisition of echo-dephased projections along all three physical axes. Dephasing is preferably set to 4π within each projection ensuring that background tissues do not contribute to signal formation and thus appear heavily suppressed. However, within the close vicinity of the paramagnetic marker, local gradient fields compensate for the intrinsic dephasing to form an echo. Successful localization of the paramagnetic marker with PRIDE is demonstrated in vitro and in vivo in the presence of different types of off-resonance (air/tissue interfaces, main magnetic field inhomogeneities, etc). In order to utilize the PRIDE sequence for vascular interventional applications, it was interleaved with balanced steady-state free precession (bSSFP) to provide positional updates to the imaged slice using a dedicated real-time feedback link. Active slice positioning (ASP) with PRIDE is demonstrated in vitro, requiring approximately 20 ms for the positional update to the imaging sequence, comparable to existing active tracking methods.

Details

einblenden:
ausblenden:
Sprache(n):
 Datum: 2009-10
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: URI: http://onlinelibrary.wiley.com/doi/10.1002/mrm.22080/pdf
DOI: 10.1002/mrm.22080
BibTex Citekey: PatilBS2009
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Magnetic Resonance in Medicine
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: - Band / Heft: 62 (4) Artikelnummer: - Start- / Endseite: 935 - 942 Identifikator: -