  Design and Evaluation of an RF Front-End for 9.4 T Human MRI

Shajan, G., Hoffmann, J., Budde, J., Adriany G, Ugurbil, K., & Pohmann, R. (2011). Design and Evaluation of an RF Front-End for 9.4 T Human MRI. Magnetic Resonance in Medicine, 66(2), 594–602. doi:10.1002/mrm.22808.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Shajan, G1, Author           
Hoffmann, J1, Author           
Budde, J1, Author           
Adriany G, Ugurbil, K, Author
Pohmann, R1, Author           
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Bmath image profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain.

Details

show
hide
Language(s):
 Dates: 2011-08
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 66 (2) Sequence Number: - Start / End Page: 594–602 Identifier: -