  MRI-Based Attenuation Correction for Whole-Body PET/MRI: Quantitative Evaluation of Segmentation- and Atlas-Based Methods

Hofmann, M., Bezrukov, I., Mantlik, F., Aschoff P, Steinke, F., Beyer T, Pichler, B., & Schölkopf, B. (2011). MRI-Based Attenuation Correction for Whole-Body PET/MRI: Quantitative Evaluation of Segmentation- and Atlas-Based Methods. Journal of Nuclear Medicine, 52(9), 1392-1399. doi:10.2967/jnumed.110.078949.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Hofmann, M1, 著者           
Bezrukov, I1, 著者           
Mantlik, F1, 著者           
Aschoff P, Steinke, F1, 著者           
Beyer T, Pichler, BJ, 著者
Schölkopf, B1, 著者           
所属:
1Department Empirical Inference, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497795              

内容説明

表示:
非表示:
キーワード: -
 要旨: PET/MRI is an emerging dual-modality imaging technology that requires new approaches to PET attenuation correction (AC). We assessed 2 algorithms for whole-body MRI-based AC (MRAC): a basic MR image segmentation algorithm and a method based on atlas registration and pattern recognition (ATPR). METHODS: Eleven patients each underwent a whole-body PET/CT study and a separate multibed whole-body MRI study. The MR image segmentation algorithm uses a combination of image thresholds, Dixon fat-water segmentation, and component analysis to detect the lungs. MR images are segmented into 5 tissue classes (not including bone), and each class is assigned a default linear attenuation value. The ATPR algorithm uses a database of previously aligned pairs of MRI/CT image volumes. For each patient, these pairs are registered to the patient MRI volume, and machine-learning techniques are used to predict attenuation values on a continuous scale. MRAC methods are compared via the quantitative analysis of AC PET images using volumes of interest in normal organs and on lesions. We assume the PET/CT values after CT-based AC to be the reference standard. RESULTS: In regions of normal physiologic uptake, the average error of the mean standardized uptake value was 14.1 ± 10.2 and 7.7 ± 8.4 for the segmentation and the ATPR methods, respectively. Lesion-based errors were 7.5 ± 7.9 for the segmentation method and 5.7 ± 4.7 for the ATPR method. CONCLUSION: The MRAC method using ATPR provided better overall PET quantification accuracy than the basic MR image segmentation approach. This better quantification was due to the significantly reduced volume of errors made regarding volumes of interest within or near bones and the slightly reduced volume of errors made regarding areas outside the lungs.

資料詳細

表示:
非表示:
言語:
 日付: 2011-09
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): URI: http://jnm.snmjournals.org/content/52/9/1392.long
DOI: 10.2967/jnumed.110.078949
BibTex参照ID: HofmannBMASBPS2011
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Journal of Nuclear Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: -
ページ: - 巻号: 52 (9) 通巻号: - 開始・終了ページ: 1392 - 1399 識別子(ISBN, ISSN, DOIなど): -