  Effect of MR Contrast Agents on Quantitative Accuracy of PET in Combined Whole-Body PET/MR Imaging

Lois, C., Bezrukov, I., Schmidt H, Schwenzer N, Werner MK, Pichler BJ, Kupferschläger, J., & Beyer, T. (2011). Effect of MR Contrast Agents on Quantitative Accuracy of PET in Combined Whole-Body PET/MR Imaging. Talk presented at 2011 IEEE Nuclear Science Symposium, Medical Imaging Conference (NSS-MIC 2011). Valencia, Spain.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Lois, C, Author
Bezrukov, I1, Author           
Schmidt H, Schwenzer N, Werner MK, Pichler BJ, Kupferschläger, J, Author
Beyer, T, Author
Affiliations:
1Department Empirical Inference, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497795              

Content

show
hide
Free keywords: -
 Abstract: Combined whole-body PET/MR systems are being tested in clinical practice today. Integrated imaging protocols entail the use of MR contrast agents (MRCA) that could bias PET attenuation correction. In this work, we assess the effect of MRCA in PET/MR imaging. We analyze the effect of oral and intravenous MRCA on PET activity after attenuation correction. We conclude that in clinical scenarios, MRCA are not expected to lead to significant attenuation of PET signals, and that attenuation maps are not biased after the ingestion of adequate oral contrasts.

Details

show
hide
Language(s):
 Dates: 2011-10
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: URI: http://www.nss-mic.org/2011/program/ListProgramDB.asp?session=MIC3
BibTex Citekey: LoisBSSWPKB2011
 Degree: -

Event

show
hide
Title: 2011 IEEE Nuclear Science Symposium, Medical Imaging Conference (NSS-MIC 2011)
Place of Event: Valencia, Spain
Start-/End Date: -

Legal Case

show

Project information

show

Source

show