English
 
Help Privacy Policy Disclaimer
  Advanced SearchBrowse

Item

ITEM ACTIONSEXPORT
  Quantitative in vivo diffusion imaging of cartilage using double echo steady-state free precession

Bieri, O., Ganter, C., & Scheffler, K. (2012). Quantitative in vivo diffusion imaging of cartilage using double echo steady-state free precession. Magnetic Resonance in Medicine, 68(3), 720–729. doi:10.1002/mrm.23275.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Bieri, O, Author
Ganter, C, Author
Scheffler, K1, Author           
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: Single-shot echo-planar imaging techniques are commonly used for diffusion-weighted imaging (DWI) but offer rather poor spatial resolution and field-of-view coverage for species with short T2. In contrast, steady-state free precession (SSFP) has shown promising results for DWI of the musculoskeletal system, but quantification is generally hampered by its prominent sensitivity on relaxation times. In this work, a new and truly diffusion-weighted (that is relaxation time independent) SSFP DWI technique is introduced using a double-echo steady-state approach. Within this framework (and this is in contrast to common SSFP DWI techniques using SSFP-Echo) both primary echo paths of nonbalanced SSFP are acquired, namely the FID and the Echo. Simulations and in vitro measurements reveal that the ratio of the Echo/FID signal ratios of two double-echo steady-state scans acquired with and without diffusion sensitizing dephasing moments provides a highly relaxation independent quantity for diffusion quantification. As a result, relaxation-independent high-resolution (0.4 × 0.4 − 0.6 × 0.6 mm2 in-plane resolution) quantitative in vivo SSFP DWI is demonstrated for human articular cartilage using diffusion-weighted double-echo steady-state scans in the knee and ankle joint at 3.0 T. The derived diffusion coefficients for cartilage (D ∼ 1.0–1.5 μm2/ms) and synovial fluid (D ∼ 2.6 μm2/ms) are in agreement with previous work.

Details

show
hide
Language(s):
 Dates: 2012-09
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: URI: http://onlinelibrary.wiley.com/doi/10.1002/mrm.23275/pdf
DOI: 10.1002/mrm.23275
BibTex Citekey: BieriGS2012_3
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Magnetic Resonance in Medicine
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 68 (3) Sequence Number: - Start / End Page: 720–729 Identifier: -