  Comprehensive inter-laboratory calibration of reference materials for δ 18O versus VSMOW using various on-line high-temperature conversion techniques

Brand, W. A., Coplen, T. B., Aerts-Bijma, A. T., Bohlke, J. K., Gehre, M., Geilmann, H., et al. (2009). Comprehensive inter-laboratory calibration of reference materials for δ 18O versus VSMOW using various on-line high-temperature conversion techniques. Rapid Communications in Mass Spectrometry, 23(7), 999-1019. doi:10.1002/rcm.3958.

Item is

Files

show Files
hide Files
:
BGC1226.pdf (Publisher version), 2MB
 
File Permalink:
-
Name:
BGC1226.pdf
Description:
-
OA-Status:
Visibility:
Restricted (Max Planck Institute for Biogeochemistry, MJBK; )
MIME-Type / Checksum:
application/octet-stream
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
License:
-

Locators

show

Creators

show
hide
 Creators:
Brand, Willi A.1, Author           
Coplen, T. B., Author
Aerts-Bijma, A. T., Author
Bohlke, J. K., Author
Gehre, M., Author
Geilmann, Heike1, Author           
Groning, M., Author
Jansen, H. G., Author
Meijer, H. A. J., Author
Mroczkowski, S. J., Author
Qi, H. P., Author
Soergel, K., Author
Stuart-Williams, H., Author
Weise, S. M., Author
Werner, R. A., Author
Affiliations:
1Service Facility Stable Isotope, Dr. W. A. Brand, Max Planck Institute for Biogeochemistry, Max Planck Society, ou_1497772              

Content

show
hide
Free keywords: Isotope ratio measurement Mass-spectrometry Cellulose synthesis Organic-substances Oxygen isotopes Water samples Exchange o-18 Fractionation Delta-c-13
 Abstract: Internationally distributed organic and inorganic oxygen isotopic reference materials have been calibrated by six laboratories carrying out more than 5300 measurements using a variety of high-temperature conversion techniques (HTC)a in an evaluation sponsored by the International Union of Pure and Applied Chemistry GUPAC). To aid in the calibration of these reference materials, which span more than 125 parts per thousand, an artificially enriched reference water (delta O-18 of +78.91 parts per thousand) and two barium sulfates (one depleted and one enriched in 180) were prepared and calibrated relative to VSMOW2(b) and SLAP reference waters. These materials were used to calibrate the other isotopic reference materials in this study, which yielded: [GRAPHICS] The seemingly large estimated combined uncertainties arise from differences in instrumentation and methodology and difficulty in accounting for all measurement bias. They are composed of the 3-fold standard errors directly calculated from the measurements and provision for systematic errors discussed in this paper. A primary conclusion of this study is that nitrate samples analyzed for delta O-18 should be analyzed with internationally distributed isotopic nitrates, and likewise for sulfates and organics. Authors reporting relative differences of oxygen-isotope ratios (delta O-18) of nitrates, sulfates, or organic material should explicitly state in their reports the 3180 values of two or more internationally distributed nitrates (USGS34, IAEA-NO-3, and USGS35), sulfates (IAEA-SO-5, IAEA-SO-6, and NBS 127), or organic material (IAEA-601 benzoic acid, IAEA-602 benzoic acid, and IAEA-600 caffeine), as appropriate to the material being analyzed, had these reference materials been analyzed with unknowns. This procedure ensures that readers will be able to normalize the delta O-18 values at a later time should it become necessary. The high-temperature reduction technique for analyzing delta O-18 and delta H-2 is not as widely applicable as the well-established combustion technique for carbon and nitrogen stable isotope determination. To obtain the most reliable stable isotope data, materials should be treated in an identical fashion; within the same sequence of analyses, samples should be compared with working reference materials that are as similar in nature and in isotopic composition as feasible. Copyright (C) 2009 John Wiley & Sons, Ltd. [References: 69]

Details

show
hide
Language(s):
 Dates: 2009
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: DOI: 10.1002/rcm.3958
Other: BGC1226
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Rapid Communications in Mass Spectrometry
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: New York, NY : John Wiley & Sons
Pages: - Volume / Issue: 23 (7) Sequence Number: - Start / End Page: 999 - 1019 Identifier: CoNE: https://pure.mpg.de/cone/journals/resource/954925574961
ISSN: 0951-4198