Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

DATENSATZ AKTIONENEXPORT
  High-Resolution Mass Spectrometry: The Trap Design and Detection System of Pentatrap and New Q-Values for Neutrino Studies

Roux, C.-E. (2012). High-Resolution Mass Spectrometry: The Trap Design and Detection System of Pentatrap and New Q-Values for Neutrino Studies. PhD Thesis, Ruprecht-Karls Universität, Heidelberg.

Item is

Dateien

einblenden: Dateien
ausblenden: Dateien
:
PhD_ChristianRoux.pdf (beliebiger Volltext), 16MB
Name:
PhD_ChristianRoux.pdf
Beschreibung:
-
OA-Status:
Sichtbarkeit:
Öffentlich
MIME-Typ / Prüfsumme:
application/pdf / [MD5]
Technische Metadaten:
Copyright Datum:
-
Copyright Info:
-
Lizenz:
-

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Roux, Christian-Eric1, Autor           
Quint, Wolfgang2, Gutachter
Blaum, Klaus, Gutachter
Affiliations:
1Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              
2GSI Helmholtzzentrum für Schwerionenforschung, Darmstadt, ou_persistent22              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: The novel Penning-trap mass spectrometer Pentatrap aims for mass-ratio measurements of highly-charged ions with a relative precision of a few parts in 1012. As the key part, an innovative trap assembly was designed. It consists of five cylindrical traps allowing for simultaneous measurements of two ion species and for continuous monitoring of the trapping conditions. This promises a substantial reduction of systematic errors. Moreover, in the course of this thesis a detection system was developed and characterized, which will enable fast and accurate measurements through its single-ion sensitivity and high signal-to-noise ratio. In neutrino physics, the detection of neutrinoless double-electron capture would unambiguously prove the Majorana nature of neutrinos. In search for a suitable nuclide for the observation of this process, precise information about the total decay energy (Q-value) is needed in order to find a transition with a resonant enhancement of the decay rate. The mass ratios of 152Sm/ 152Gd and 164Dy/ 164Er were measured to a relative uncertainty of ~ 10−9 at Shiptrap. This presents the first directly determined Q-values for the corresponding transitions.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2012-11-05
 Publikationsstatus: Angenommen
 Seiten: VII, 182 S. : Ill., graph. Darst.
 Ort, Verlag, Ausgabe: Heidelberg : Ruprecht-Karls Universität
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: -
 Art des Abschluß: Doktorarbeit

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle