  Performance of a short “magnetic bottle” electron spectrometer

Mucke, M., Förstel, M., Lischke, T., Arion, T., Bradshaw, A. M., & Hergenhahn, U. (2012). Performance of a short “magnetic bottle” electron spectrometer. Review of Scientific Instruments, 83(6): 063106. doi:10.1063/1.4729256.

Item is

Files

show Files
hide Files
:
magb-edoc.pdf (Any fulltext), 259KB
 
File Permalink:
-
Name:
magb-edoc.pdf
Description:
-
OA-Status:
Visibility:
Private
MIME-Type / Checksum:
application/pdf
Technical Metadata:
Copyright Date:
2012
Copyright Info:
AIP
License:
-
:
1.4729256.pdf (Publisher version), 744KB
Name:
1.4729256.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
2012
Copyright Info:
AIP
License:
OA Nationallizenz

Locators

show

Creators

show
hide
 Creators:
Mucke, Melanie1, Author
Förstel, Marko1, 2, Author
Lischke, Toralf1, 3, Author           
Arion, Tiberiu1, Author
Bradshaw, Alex M.1, 3, Author           
Hergenhahn, Uwe4, Author
Affiliations:
1Max-Planck-Institut für Plasmaphysik, EURATOM Association, Boltzmannstr. 2, 85748 Garching, Germany , ou_persistent22              
2Max-Planck-Institut für Kernphysik, Saupfercheckweg 1, 69117 Heidelberg, Germany , ou_persistent22              
3Molecular Physics, Fritz Haber Institute, Max Planck Society, ou_634545              
4Max-Planck-Institut für Plasmaphysik, EURATOM Association, Teilinstitut Greifswald, Wendelsteinstr. 1, 17491 Greifswald, Germany, ou_persistent22              

Content

show
hide
Free keywords: -
 Abstract: In this article, a newly constructed electron spectrometer of the magnetic bottle type is described. The instrument is part of an apparatus for measuring the electron spectra of free clusters using synchrotron radiation. Argon and helium outer valence photoelectron spectra have been recorded in order to investigate the characteristic features of the spectrometer. The energy resolution (E/ΔE) has been found to be ∼30. Using electrostatic retardation of the electrons, it can be increased to at least 110. The transmission as a function of kinetic energy is flat, and is not impaired much by retardation with up to 80% of the initial kinetic energy. We have measured a detection efficiency of most probably 0.6 −0.1+0.05, but at least of 0.4. Results from testing the alignment of the magnet, and from trajectory simulations, are also discussed.

Details

show
hide
Language(s): eng - English
 Dates: 2012-04-192012-05-252012-06-142012-06
 Publication Status: Issued
 Pages: 9
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Review of Scientific Instruments
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Woodbury, N.Y. [etc.] : American Institute of Physics
Pages: - Volume / Issue: 83 (6) Sequence Number: 063106 Start / End Page: - Identifier: ISSN: 0034-6748
CoNE: https://pure.mpg.de/cone/journals/resource/991042742033452