  Endotoxins in ophthalmic viscosurgical devices

Dick, H. B., Augustin, A. J., Pakula, T., & Pfeiffer, N. (2003). Endotoxins in ophthalmic viscosurgical devices. European Journal of Ophthalmology, 13(2), 176-184.

Item is

Externe Referenzen

einblenden:

Urheber

einblenden:
ausblenden:
 Urheber:
Dick, H. B., Autor
Augustin, A. J., Autor
Pakula, Tadeusz1, Autor           
Pfeiffer, N., Autor
Affiliations:
1MPI for Polymer Research, Max Planck Society, ou_1309545              

Inhalt

einblenden:
ausblenden:
Schlagwörter: ophthalmic viscosurgical devices; endotoxins; limulus amebocyte lysate assay; inflammatory reaction
 Zusammenfassung: PURPOSE. To measure the endotoxin concentration (EC) of 25 commercially available, hyaluronic acid- and hydroxypropylmethylcellulose-based (HPMC) ophthalmic viscosurgical devices (OVDs). METHODS. The in vitro Limulus amebocyte lysate (LAL) assay, which indicates the presence of endotoxins originating from gram-negative bacteria, was used to determine the EC. The procedure was performed according to the European Pharmacopoeia/USP. EC including duplicate determinations, negative controls, dilution series with control standard endotoxin, dilution series with sample extract and positive sample control. RESULTS. 16 OVDs (Amvisc(R), Amvisc(R) Plus, Biolon(R), Coatel(R), Healon(R), Healon(R) GV, Healon(R)5, HPMC Ophtal(R) L, Microvisc(R), Microvisc(R) Plus, Ocucoat(R), Provisc(R), Rayvisc(R), Viscoat(R), Visco Shield(R) 2%, Visko(R) 1.4%) had an EC under 1.2 endotoxin units/mL, five (Adatocel(R), HPMC Ophtal(R) H, LA Gel(R), Viscorneal(R), Viscorneal(R) Plus) had an EC greater than or equal to 1.2 and less than or equal to 24 EU/ml, and four (Biocorneal(R), Dispasan(R) also named Ophthalin, Dispasan(R) Plus, Visko(R), 1%) had an EC of > 24 EU/ml. DISCUSSION. To avoid viscoelastic- related inflammatory or immunological reactions, the use of pure OVDs is recommended, especially for surgical procedures with an inherent possibility of leaving viscoelastic remnants in the eye (e.g., cataract surgery, visco-canalostomy or penetrating keratoplasty).

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2003-03
 Publikationsstatus: Erschienen
 Seiten: -
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: -
 Identifikatoren: eDoc: 28377
ISI: 000181837900009
Anderer: P-03-123
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: European Journal of Ophthalmology
  Alternativer Titel : Eur. J. Ophthalmol.
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: -
Seiten: - Band / Heft: 13 (2) Artikelnummer: - Start- / Endseite: 176 - 184 Identifikator: ISSN: 1120-6721