  TITAN’s digital RFQ ion beam cooler and buncher,operation and performance

Brunner, T., Smith, M., Brodeur, M., Ettenauer, S., Gallant, A., Simon, V. V., et al. (2012). TITAN’s digital RFQ ion beam cooler and buncher,operation and performance. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 676, 32-43. doi:10.1016/j.nima.2012.02.004.

Item is

Files

show Files

Locators

show
hide
Locator:
http://dx.doi.org/10.1016/j.nima.2012.02.004 (Publisher version)
Description:
-
OA-Status:

Creators

show
hide
 Creators:
Brunner, T.1, 2, Author
Smith, M.J.1, 3, Author
Brodeur, M.1, 3, Author
Ettenauer, S.1, 3, Author
Gallant, A.T.1, 3, Author
Simon, Vanessa V.4, Author           
Chaudhuri, A.1, Author
Lapierre, A.1, Author
Mané, E.1, Author
Ringle, R.1, Author
Simon, M.C.1, Author
Vaz, J.A.1, Author
Delheij, P.3, Author
Good, M.1, Author
Pearson, M.R.1, Author
Dilling, Jens1, 3, Author
Affiliations:
1TRIUMF, 4004 Westbrook Mall,Vancouver,BC,CanadaV6T2A3, ou_persistent22              
2Physik Department E12,Technische Universität München, James Franck Straße,D-85748 Garching, Germany, ou_persistent22              
3Department of Physics and Astronomy,University of British Columbia, 6224 Agriculture Road,Vancouver,BC,Canada V6T1Z1, ou_persistent22              
4Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Content

show
hide
Free keywords: -
 Abstract: We present a description of the Radio Frequency Quadrupole (RFQ) ion trap built as part of the TITAN facility. It consists of a gas-filled, segmented, linear Paul trap and is the first stage of the TITAN setup with the purpose of cooling and bunching radioactive ion beams delivered from ISAC-TRIUMF. This is the first such device to be driven digitally, i.e., using a high voltage (View the MathML source), wide bandwidth (View the MathML source) square-wave as compared to the typical sinusoidal wave form. Results from the commissioning of the device as well as systematic studies with stable and radioactive ions are presented including efficiency measurements with stable 133Cs and radioactive 124,126Cs. A novel and unique mode of operation of this device is also demonstrated where the cooled ion bunches are extracted in reverse mode, i.e., in the same direction as previously injected.

Details

show
hide
Language(s): eng - English
 Dates: 2012-02-10
 Publication Status: Published online
 Pages: 12
 Publishing info: -
 Table of Contents: -
 Rev. Type: Peer
 Identifiers: DOI: 10.1016/j.nima.2012.02.004
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
  Other : Nucl. Instrum. Methods Phys. Res. A
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: North-Holland
Pages: - Volume / Issue: 676 Sequence Number: - Start / End Page: 32 - 43 Identifier: ISSN: 0168-9002
CoNE: https://pure.mpg.de/cone/journals/resource/954925485716