日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

登録内容を編集ファイル形式で保存
 
 
ダウンロード電子メール
  Correction of inter-scan motion artifacts in quantitative R1 mapping by accounting for receive coil sensitivity effects

Papp, D., Callaghan, M. F., Meyer, H., Buckley, C., & Weiskopf, N. (2016). Correction of inter-scan motion artifacts in quantitative R1 mapping by accounting for receive coil sensitivity effects. Magnetic Resonance in Medicine, 76(5), 1478-1485. doi:10.1002/mrm.26058.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル
非表示: ファイル
:
Papp_2016.pdf (出版社版), 522KB
ファイルのパーマリンク:
https://hdl.handle.net/21.11116/0000-0001-D74E-6
ファイル名:
Papp_2016.pdf
説明:
-
OA-Status:
閲覧制限:
公開
MIMEタイプ / チェックサム:
application/pdf / [MD5]
技術的なメタデータ:
著作権日付:
-
著作権情報:
-
CCライセンス:
-

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Papp, Daniel1, 著者
Callaghan, Martina F.1, 著者
Meyer, Heiko2, 著者
Buckley, Craig3, 著者
Weiskopf, Nikolaus1, 4, 著者           
所属:
1Wellcome Trust Centre for Neuroimaging, University College London, United Kingdom, ou_persistent22              
2Siemens Sector Healthcare, Erlangen, Germany, ou_persistent22              
3Siemens Healthcare, Camberley, United Kingdom, ou_persistent22              
4Department Neurophysics (Weiskopf), MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_2205649              

内容説明

表示:
非表示:
キーワード: MPM; R1 mapping; VFA; Motion correction; Quantitative MRI; Receive sensitivity
 要旨: PURPOSE:

Inter-scan motion causes differential receive field modulation between scans, leading to errors when they are combined to quantify MRI parameters. We present a robust and efficient method that accounts for inter-scan motion by removing this modulation before parameter quantification.
THEORY AND METHODS:

Five participants moved between two high-resolution structural scans acquired with different flip angles. Before each high-resolution scan, the effective relative sensitivity of the receive head coil was estimated by combining two rapid low-resolution scans acquired receiving on each of the body and head coils. All data were co-registered and sensitivity variations were removed from the high-resolution scans by division with the effective relative sensitivity. R1 maps with and without this correction were calculated and compared against reference maps unaffected by inter-scan motion.
RESULTS:

Even after coregistration, inter-scan motion significantly biased the R1 maps, leading to spurious variation in R1 in brain tissue and deviations with respect to a no-motion reference. The proposed correction scheme reduced the error to within the typical scan-rescan error observed in datasets unaffected by motion.
CONCLUSION:

Inter-scan motion negatively impacts the accuracy and precision of R1 mapping. We present a validated correction method that accounts for position-specific receive field modulation. Magn Reson Med, 2015. © 2015 The Authors. Magnetic Resonance in Medicine published by Wiley Periodicals, Inc. on behalf of International Society for Magnetic Resonance in Medicine. This is an open access article under the terms of the Creative Commons Attribution License, which permits use, distribution and reproduction in any medium, provided the original work is properly cited.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2015-10-062015-06-252015-10-262015-11-262016-11-01
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.26058
PMID: 26608936
PMC: PMC5082493
その他: Epub 2015
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: New York : Wiley-Liss
ページ: - 巻号: 76 (5) 通巻号: - 開始・終了ページ: 1478 - 1485 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149