  Calibration of sealed HCl cells used for TCCON instrumental line shape monitoring

Hase, F., Drouin, B. J., Roehl, C. M., Toon, G. C., Wennberg, P. O., Wunch, D., et al. (2013). Calibration of sealed HCl cells used for TCCON instrumental line shape monitoring. Atmospheric Measurement Techniques, 6, 3527-3537. doi:10.5194/amt-6-3527-2013.

Item is

Files

show Files
hide Files
:
BGC1883D.pdf (Publisher version), 7MB
Name:
BGC1883D.pdf
Description:
Discussion paper
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-
:
BGC1883.pdf (Publisher version), 7MB
Name:
BGC1883.pdf
Description:
-
OA-Status:
Visibility:
Public
MIME-Type / Checksum:
application/pdf / [MD5]
Technical Metadata:
Copyright Date:
-
Copyright Info:
-

Locators

show
hide
Locator:
http://dx.doi.org/10.5194/amt-6-3527-2013 (Publisher version)
Description:
OA
OA-Status:

Creators

show
hide
 Creators:
Hase, F., Author
Drouin, B. J., Author
Roehl, C. M., Author
Toon, G. C., Author
Wennberg, P. O., Author
Wunch, D., Author
Blumenstock, T., Author
Desmet, F., Author
Feist, Dietrich G.1, Author           
Heikkinen, P., Author
De Maziere, M., Author
Rettinger, M., Author
Robinson, J., Author
Schneider, M., Author
Sherlock, V., Author
Sussmann, R., Author
Te, Y., Author
Warneke, T., Author
Weinzierl, C., Author
Affiliations:
1Atmospheric Remote Sensing Group, Dr. D. Feist, Department Biogeochemical Systems, Prof. M. Heimann, Max Planck Institute for Biogeochemistry, Max Planck Society, ou_1497783              

Content

show
hide
Free keywords: -
 Abstract: The TCCON (Total Carbon Column Observing Network) FTIR network provides highly accurate observations of greenhouse gases column-averaged dry-air mole fractions. As an important component of TCCON quality assurance measures, sealed cells 5 filled with approx. 5mbar of HCl are used for instrumental line shape (ILS) monitoring at all TCCON sites. Here, we introduce a calibration procedure for the HCl cells which applies a refillable, pressure-monitored reference cell filled with C2H2. Using this method, we identify residual variations of HCl purity between the TCCON cells as a non-negligible disturbance.

Details

show
hide
Language(s):
 Dates: 2013-07-232013-12-162013
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: Other: BGC1883
DOI: 10.5194/amt-6-3527-2013
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Atmospheric Measurement Techniques
  Abbreviation : AMT
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: Göttingen : European Geosciences Union, Copernicus
Pages: - Volume / Issue: 6 Sequence Number: - Start / End Page: 3527 - 3537 Identifier: ISSN: 1867-1381
CoNE: https://pure.mpg.de/cone/journals/resource/1867-1381