  True real-time cardiac MRI in free breathing without ECG synchronization using a novel sequence with radial k-space sampling and balanced SSFP contrast mode.

Bauer, R. W., Radtke, I., Block, K. T., Larson, M. C., Kerl, J. M., Hammerstingl, R., Graf, T. G., Vogl, T. J., & Zhang, S. (2013). True real-time cardiac MRI in free breathing without ECG synchronization using a novel sequence with radial k-space sampling and balanced SSFP contrast mode. The International Journal of Cardiovascular Imaging, 29(5), 1059-1067. doi:10.1007/s10554-013-0183-0.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル
非表示: ファイル
:
1833516.pdf (出版社版), 435KB
 
ファイルのパーマリンク:
-
ファイル名:
1833516.pdf
説明:
-
OA-Status:
閲覧制限:
制限付き (UNKNOWN id 303; )
MIMEタイプ / チェックサム:
application/pdf
技術的なメタデータ:
著作権日付:
-
著作権情報:
-
CCライセンス:
-

作成者

表示:
非表示:
 作成者:
Bauer, R. W., 著者
Radtke, I., 著者
Block, K. T., 著者
Larson, M. C., 著者
Kerl, J. M., 著者
Hammerstingl, R., 著者
Graf, T. G., 著者
Vogl, T. J., 著者
Zhang, S.1, 著者           
所属:
1Biomedical NMR Research GmbH, MPI for biophysical chemistry, Max Planck Society, ou_578634              

内容説明

表示:
非表示:
キーワード: Real-time MRI; Cardiac MRI; Radial sampling; Cardiac function
 要旨: We investigated a novel sequence with radial k-space sampling, gridding and sliding window reconstruction with bSSFP contrast that allows for true real-time functional cardiac evaluation independent from respiration and ECG triggering. 12 healthy volunteers underwent 1.5 T cardiac MRI. Single-shot short axis views were acquired with a) standard retrospectively ECG-gated segmented breath-hold (bh) bSSFP and with the real-time radial bSSFP sequence with a nominal temporal resolution of b) 16 fps (frames per second) and c) 40 fps. Radial bSSFP were acquired during free breathing without ECG synchronization. Left ventricular functional parameters (EDV, ESV, SV and EF) were compared and quality of wall motion depiction was assessed. Contrast-to-noise-ratio (CNR) of myocardium/blood pool in the left ventricle was calculated. EF showed excellent correlation (Bland-Altman r = 0.99; Lin rho = 0.91) between bh-bSSFP (65 %) and 40 fps radial (64 %) and moderate correlation (r = 0.84, rho = 0.20) with 16 fps radial bSSFP (56 %). While EDV was in good agreement for all three sequences, ESV was significantly overestimated with 16 fps radial bSSFP. Despite lower CNR, image quality for wall motion assessment was rated significantly better for 40 fps compared to 16 fps radial bSSFP due to the faster temporal resolution. Left ventricular functional analysis with fast true real-time radial bSSFP is in good agreement with standard ECG-gated bh-bSSFP. The independency from ECG synchronization and breathing promises a robust method for patients with impaired cardiopulmonary status in whom breath-hold and good quality ECG cannot be achieved.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2013-01-192013-06
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): DOI: 10.1007/s10554-013-0183-0
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: The International Journal of Cardiovascular Imaging
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: -
ページ: - 巻号: 29 (5) 通巻号: - 開始・終了ページ: 1059 - 1067 識別子(ISBN, ISSN, DOIなど): -