日本語
 
Help Privacy Policy ポリシー/免責事項
  詳細検索ブラウズ

アイテム詳細

登録内容を編集ファイル形式で保存
 
 
ダウンロード電子メール
  Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T with minimal signal loss

Henning, A., Fuchs A, Murdoch, J., & Boesiger, P. (2009). Slice-selective FID acquisition, localized by outer volume suppression (FIDLOVS) for 1H-MRSI of the human brain at 7 T with minimal signal loss. NMR in Biomedicine, 122(7), 683–696. doi:10.1002/nbm.1366.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Henning, A1, 著者           
Fuchs A, Murdoch, JB, 著者
Boesiger, P, 著者
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

内容説明

表示:
非表示:
キーワード: -
 要旨: In comparison to 1.5 and 3 T, MR spectroscopic imaging at 7 T benefits from signal-to-noise ratio (SNR) gain and increased spectral resolution and should enable mapping of a large number of metabolites at high spatial resolutions. However, to take full advantage of the ultra-high field strength, severe technical challenges, e.g. related to very short T2 relaxation times and strict limitations on the maximum achievable B1 field strength, have to be resolved. The latter results in a considerable decrease in bandwidth for conventional amplitude modulated radio frequency pulses (RF-pulses) and thus to an undesirably large chemical-shift displacement artefact. Frequency-modulated RF-pulses can overcome this problem; but to achieve a sufficient bandwidth, long pulse durations are required that lead to undesirably long echo-times in the presence of short T2 relaxation times. In this work, a new magnetic resonance spectroscopic imaging (MRSI) localization scheme (free induction decay acquisition localized by outer volume suppression, FIDLOVS) is introduced that enables MRSI data acquisition with minimal SNR loss due to T2 relaxation and thus for the first time mapping of an extended neurochemical profile in the human brain at 7 T. To overcome the contradictory problems of short T2 relaxation times and long pulse durations, the free induction decay (FID) is directly acquired after slice-selective excitation. Localization in the second and third dimension and skull lipid suppression are based on a T1- and B1-insensitive outer volume suppression (OVS) sequence. Broadband frequency-modulated excitation and saturation pulses enable a minimization of the chemical-shift displacement artefact in the presence of strict limits on the maximum B1 field strength. The variable power RF pulses with optimized relaxation delays (VAPOR) water suppression scheme, which is interleaved with OVS pulses, eliminates modulation side bands and strong baseline distortions. Third order shimming is based on the accelerated projection-based automatic shimming routine (FASTERMAP) algorithm. The striking SNR and spectral resolution enable unambiguous quantification and mapping of 12 metabolites including glutamate (Glu), glutamine (Gln), N-acetyl-aspartatyl-glutamate (NAAG), γ-aminobutyric acid (GABA) and glutathione (GSH). The high SNR is also the basis for highly spatially resolved metabolite mapping.

資料詳細

表示:
非表示:
言語:
 日付: 2009-08
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): URI: http://onlinelibrary.wiley.com/doi/10.1002/nbm.1366/pdf
DOI: 10.1002/nbm.1366
BibTex参照ID: HenningFMB2009
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: NMR in Biomedicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: -
ページ: - 巻号: 122 (7) 通巻号: - 開始・終了ページ: 683–696 識別子(ISBN, ISSN, DOIなど): -