  A four-channel hole-slotted phased array at 7 Tesla

Lopez, M., Ehses, P., Breuer FA, Ponce IP, Gareis, D., & Jakob, P. (2010). A four-channel hole-slotted phased array at 7 Tesla. Concepts in Magnetic Resonance B: Magnetic Resonance Engineering, 37B(4), 226-236. doi:10.1002/cmr.b.20173.

Item is

Files

show Files

Locators

show

Creators

show
hide
 Creators:
Lopez, MA, Author
Ehses, P1, Author           
Breuer FA, Ponce IP, Gareis, D, Author
Jakob, PM, Author
Affiliations:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

Content

show
hide
Free keywords: -
 Abstract: In recent years, the phased array coil technology has found more and more its way in applications at high field small animal systems. However, these coil arrays are usually based on simple loops and are mostly used in rat imaging studies only. Mouse imaging studies are limited to the use of linear arrays or volume resonators. Recently, a novel surface coil design based on the hole-slot magnetron's geometry was introduced to MRI. The hole-slot magnetron is a vacuum tube which operates e.g. as a high frequency oscillator in radar applications. It has been shown that the magnetron surface coil allows for a deeper RF penetration than a conventional coil both at 1.5 and 4 Tesla. The objective of this work was to find an optimal loop coil based on the hole-slot magnetron geometry in order to build a volume phased array for cardiac imaging with improved SNR on the centre of the sample. To achieve this goal, different magnetron loops were simulated and evaluated towards their performance. Based on these results the best performing hole-slot magnetron geometries were built and compared. In addition, the magnetron loop (referred as hole-slotted coil) with the best sensitivity was compared with conventional simple loop geometries. Furthermore, a four channel hole-slotted phased array based on the magnetron's design theory was built, evaluated and compared with a conventional four channel array. The four channel hole-slotted array shows improved RF penetration depth over the four channel array with simple loop geometry.

Details

show
hide
Language(s):
 Dates: 2010-10
 Publication Status: Issued
 Pages: -
 Publishing info: -
 Table of Contents: -
 Rev. Type: -
 Identifiers: URI: http://onlinelibrary.wiley.com/doi/10.1002/cmr.b.20173/pdf
DOI: 10.1002/cmr.b.20173
BibTex Citekey: LopezEBPGJ2010
 Degree: -

Event

show

Legal Case

show

Project information

show

Source 1

show
hide
Title: Concepts in Magnetic Resonance B: Magnetic Resonance Engineering
Source Genre: Journal
 Creator(s):
Affiliations:
Publ. Info: -
Pages: - Volume / Issue: 37B (4) Sequence Number: - Start / End Page: 226 - 236 Identifier: -