  Real-time motion detection for MR spectroscopy in the spinal cord

Hock, A., Kollias Ss, Boesiger, P., & Henning, A. (2012). Real-time motion detection for MR spectroscopy in the spinal cord. Talk presented at 29th Annual Scientific Meeting ESMRMB 2012. Lisboa, Portugal.

Item is

基本情報

表示: 非表示:
資料種別: 講演

ファイル

表示: ファイル

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Hock, A, 著者
Kollias Ss, Boesiger, P, 著者
Henning, A1, 著者           
所属:
1Department High-Field Magnetic Resonance, Max Planck Institute for Biological Cybernetics, Max Planck Society, ou_1497796              

内容説明

表示:
非表示:
キーワード: -
 要旨: Purpose/Introduction: Subject motion is one of the major problems for MR spectroscopy (MRS) since it often leads to increased linewidth, frequency shifts, reduced peak areas, insufficient water suppression(1), and to a displacement of the acquisition voxel. Especially if the region of interest is small (like the spinal cord or tumors) even small movements will shift the measurement voxel out of the region resulting in inaccurate results. The detection of patient motion by observing exclusively the spectra is sometimes impossible. Therefore, in this work, 1D navigator acquisitions interleaved with the MRS measurement are used for real-time patient-motion detection in the spinal cord. Subjects and Methods: Non-water-suppressed MRS via the metabolite cycling technique at 3T (Achieva, Philips Healthcare, Best, TE/TR 30/2500ms, voxel size 1.2ml, 4*128FIDs) at the cervical level C3-4 (Fig.1) was used in the spinal cord(2). Interleaved navigators (Fig.2) were placed above the MRS-voxel (Fig.1). The software allows real-time data processing so that the navigator position evaluation can be displayed online (Fig.3). After the approval from the local ethics committee, one volunteer was scanned with (scan 23) and without (scan 1) navigator motion detection. The volunteer was asked to lie still during all measurements except of scans 3. There the volunteer was asked to slowly raise her chin in the middle of the scan of about 2cm and to return to initial position after ~30s (~15 FIDs, ~3 of MRS scan-time). Before and after each MRS scan axial T2-weighted images were acquired. If these images were shifted more than 1mm the MRS measurement was repeated. Results: The spectra of scan 12 (Fig.3) show that the addition of navigators to the MRS sequence does not negatively affect the MRS quality. In addition, compared to scans without motion (12) scan 3 shows a different spectral fingerprint (e.g. increased lipid signal) (Fig.3). Moreover, Fig.3 shows that even small movements are detectable by using navigators. Discussion/Conclusion: Intra-scan subject movement can lead to a significant distortion of spectra (Fig. 3, scan 3). However, especially in the clinical routine subject movement cannot be eliminated. Thus, motion detection is essential. Interleaved navigators allow precise real-time motion detection for MRS without the need of additional scan-time and additional hardware. It allows early intervention (e.g. asking the subject to lay still) and to reject motion corrupted FIDs. The use of additional navigators allows the detection of motion in more dimensions and offers the possibility to update the voxel position.

資料詳細

表示:
非表示:
言語:
 日付: 2012-10
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: -
 識別子(DOI, ISBNなど): URI: http://link.springer.com/content/pdf/10.10072Fs10334-012-0323-x
DOI: 10.1007/s10334-012-0323-x
BibTex参照ID: HockKBH2012
 学位: -

関連イベント

表示:
非表示:
イベント名: 29th Annual Scientific Meeting ESMRMB 2012
開催地: Lisboa, Portugal
開始日・終了日: -

訴訟

表示:

Project information

表示:

出版物

表示: