  Fast accurate MR thermometry using phase referenced asymmetric spin-echo EPI at high field

Streicher, M., Schäfer, A., Ivanov, D., Müller, D. K., Amadon, A., Reimer, E., Huber, L., Dhital, B., Rivera, D., Kögler, C., Trampel, R., Pampel, A., & Turner, R. (2014). Fast accurate MR thermometry using phase referenced asymmetric spin-echo EPI at high field. Magnetic Resonance in Medicine, 71(2), 524-533. doi:10.1002/mrm.24681.

Item is

基本情報

表示: 非表示:
資料種別: 学術論文

ファイル

表示: ファイル
非表示: ファイル
:
Streicher_2013_Fast.pdf (プレプリント), 636KB
 
ファイルのパーマリンク:
-
ファイル名:
Streicher_2013_Fast.pdf
説明:
-
OA-Status:
閲覧制限:
非公開
MIMEタイプ / チェックサム:
application/pdf
技術的なメタデータ:
著作権日付:
-
著作権情報:
-
CCライセンス:
-

関連URL

表示:

作成者

表示:
非表示:
 作成者:
Streicher, Markus1, 著者           
Schäfer, Andreas1, 著者           
Ivanov, Dimo1, 著者           
Müller, Dirk K.2, 著者           
Amadon, Alexis3, 著者
Reimer, Enrico1, 著者           
Huber, Laurentius1, 著者           
Dhital, Bibek1, 著者           
Rivera, Deborah1, 著者           
Kögler, Carsten1, 著者           
Trampel, Robert1, 著者           
Pampel, André2, 著者           
Turner, Robert1, 著者           
所属:
1Department Neurophysics, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634550              
2Methods and Development Unit Nuclear Magnetic Resonance, MPI for Human Cognitive and Brain Sciences, Max Planck Society, ou_634558              
3Neurospin, French Alternative Energies and Atomic Energy Commission (CEA), Gif-sur-Yvette, France, ou_persistent22              

内容説明

表示:
非表示:
キーワード: Referenced; MR thermometry; Fat; Spin-echo; Asymmetric; Proton resonance frequency shift
 要旨: Purpose:

A novel highly accurate method for MR thermometry, effective at high field, is introduced and validated, which corrects for slow and fast field fluctuations by means of reference images.

Methods:

An asymmetric spin-echo (ASE) EPI sequence was made frequency-selective to water or a reference substance by controlling the slice-select gradient polarity, or the durations of the excitation and refocusing RF pulses. Images were acquired pairwise, and the temperature-sensitive water images were corrected for field fluctuations using the reference images. In a phantom RF heating experiment, dissolved dimethyl sulfoxide was used as a reference substance. Temperature stability was tested in-vivo on the human brain, referenced using subcutaneous scalp fat. Water and fat phase images were acquired only 50 ms apart. Bloch simulations validated the frequency selection accuracy.

Results:

ASE imaging using simple frequency selection methods provides highly accurate referenced MR thermometry in phantoms and in-vivo at 7 Tesla. Effects of field fluctuations caused by field drift, breathing and heart beat were corrected. The technique is highly robust against B1 inhomogeneities.

Conclusion:

Frequency selection using gradient-reversal can enable fast accurate referenced in-vivo MR thermometry, assisting thermal characterization of RF coils and possibly in-vivo SAR monitoring.

資料詳細

表示:
非表示:
言語: eng - English
 日付: 2012-11-272012-08-312013-01-152013-02-252014-02
 出版の状態: 出版
 ページ: -
 出版情報: -
 目次: -
 査読: 査読あり
 識別子(DOI, ISBNなど): DOI: 10.1002/mrm.24681
PMID: 23440917
 学位: -

関連イベント

表示:

訴訟

表示:

Project information

表示:

出版物 1

表示:
非表示:
出版物名: Magnetic Resonance in Medicine
種別: 学術雑誌
 著者・編者:
所属:
出版社, 出版地: -
ページ: - 巻号: 71 (2) 通巻号: - 開始・終了ページ: 524 - 533 識別子(ISBN, ISSN, DOIなど): ISSN: 0740-3194
CoNE: https://pure.mpg.de/cone/journals/resource/954925538149