Deutsch
 
Hilfe Datenschutzhinweis Impressum
  DetailsucheBrowse

Datensatz

 
 
DownloadE-Mail
  TITAN’s digital RFQ ion beam cooler and buncher,operation and performance

Brunner, T., Smith, M., Brodeur, M., Ettenauer, S., Gallant, A., Simon, V. V., et al. (2012). TITAN’s digital RFQ ion beam cooler and buncher,operation and performance. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 676, 32-43. doi:10.1016/j.nima.2012.02.004.

Item is

Externe Referenzen

einblenden:
ausblenden:
externe Referenz:
http://dx.doi.org/10.1016/j.nima.2012.02.004 (Verlagsversion)
Beschreibung:
-
OA-Status:

Urheber

einblenden:
ausblenden:
 Urheber:
Brunner, T.1, 2, Autor
Smith, M.J.1, 3, Autor
Brodeur, M.1, 3, Autor
Ettenauer, S.1, 3, Autor
Gallant, A.T.1, 3, Autor
Simon, Vanessa V.4, Autor           
Chaudhuri, A.1, Autor
Lapierre, A.1, Autor
Mané, E.1, Autor
Ringle, R.1, Autor
Simon, M.C.1, Autor
Vaz, J.A.1, Autor
Delheij, P.3, Autor
Good, M.1, Autor
Pearson, M.R.1, Autor
Dilling, Jens1, 3, Autor
Affiliations:
1TRIUMF, 4004 Westbrook Mall,Vancouver,BC,CanadaV6T2A3, ou_persistent22              
2Physik Department E12,Technische Universität München, James Franck Straße,D-85748 Garching, Germany, ou_persistent22              
3Department of Physics and Astronomy,University of British Columbia, 6224 Agriculture Road,Vancouver,BC,Canada V6T1Z1, ou_persistent22              
4Division Prof. Dr. Klaus Blaum, MPI for Nuclear Physics, Max Planck Society, ou_904548              

Inhalt

einblenden:
ausblenden:
Schlagwörter: -
 Zusammenfassung: We present a description of the Radio Frequency Quadrupole (RFQ) ion trap built as part of the TITAN facility. It consists of a gas-filled, segmented, linear Paul trap and is the first stage of the TITAN setup with the purpose of cooling and bunching radioactive ion beams delivered from ISAC-TRIUMF. This is the first such device to be driven digitally, i.e., using a high voltage (View the MathML source), wide bandwidth (View the MathML source) square-wave as compared to the typical sinusoidal wave form. Results from the commissioning of the device as well as systematic studies with stable and radioactive ions are presented including efficiency measurements with stable 133Cs and radioactive 124,126Cs. A novel and unique mode of operation of this device is also demonstrated where the cooled ion bunches are extracted in reverse mode, i.e., in the same direction as previously injected.

Details

einblenden:
ausblenden:
Sprache(n): eng - English
 Datum: 2012-02-10
 Publikationsstatus: Online veröffentlicht
 Seiten: 12
 Ort, Verlag, Ausgabe: -
 Inhaltsverzeichnis: -
 Art der Begutachtung: Expertenbegutachtung
 Identifikatoren: DOI: 10.1016/j.nima.2012.02.004
 Art des Abschluß: -

Veranstaltung

einblenden:

Entscheidung

einblenden:

Projektinformation

einblenden:

Quelle 1

einblenden:
ausblenden:
Titel: Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
  Andere : Nucl. Instrum. Methods Phys. Res. A
Genre der Quelle: Zeitschrift
 Urheber:
Affiliations:
Ort, Verlag, Ausgabe: North-Holland
Seiten: - Band / Heft: 676 Artikelnummer: - Start- / Endseite: 32 - 43 Identifikator: ISSN: 0168-9002
CoNE: https://pure.mpg.de/cone/journals/resource/954925485716